Parr Instrument 6400 User Manual
Page 38

6400
Calorimeter Instruction Manual
5-6
Print Error Messages:
When turned ON, all error messages will be printed on the printer as well as displayed
on the screen. When turned OFF, messages will only display on the screen.
Language:
Steps the Calorimeter through the installed operating languages.
Program Information and
Control Menu
Date:
Displays current date and accesses
sub-menu on which date is set in
(YY/MM/DD) format.
Time:
Displays current time and accesses
sub-menu on which time is set in
(HH:MM) format.
Software and Hardware Info:
This screen displays important
information such as the main
software version, I/O board
information, CPU information, and
Controller IP address assigned by
the network DHCP server.
Settings Protect:
Provides protection for the program
options and settings on the menus.
If this is turned ON, the user will be
warned that enumeration keys are
locked when a key is pressed.
Enumeration Keys either toggle a
value (ON / OFF) or select from a
predefined list. This feature is used
primarily to protect the instrument settings from accidental changes if one were to
inadvertently touch or bump up against the touchscreen.
User/Factory Settings:
This key leads to a sub-menu that allows the user to save or recall user defined
instrument settings. Additionally, factory preinstalled settings supporting different
bombs or special operating modes can also be recalled.