beautypg.com

Tables – Parr Instrument 6400 User Manual

Page 100

background image

6400

Calorimeter Instruction Manual

14-1

TABLES

Table 2 - Default Settings

Calorimeter Operation

Operating Mode

Determination

Bomb Installed/EE

1/940

Heater and Pump

OFF


Operating Controls

Method of Operation

Dynamic

Reporting Units

btu/lb

Use Spiking Correction

OFF

“OTHER” Multiplier

4.1868

Calibrate Touchscreen

LCD Backlight Timeout(s)

1200 S

LCD Contrast

30%

Print Error Messages

ON

Language English

Spike Controls

Use Spiking

OFF

Heat of Combustion of Spike

6318.4

Use Fixed Spike

OFF

Weight of Fixed Spike

0.0

Prompt for Spike before Weight

OFF


Bomb Rinse Tank Control

Report Rinse Tank Empty

ON

Rinse Tank Capacity

150

# Rinses Left

150

Reset Rinse Tank Counter

Rinse Time

25

Rinse Flush Time

20

Clear Time

100

# of Rinse Cycles

3


Program Information and Controls

Date XX/XX/XXXX
Time XX:XX
Software and Hardware Info

Settings Protect

OFF

User/Factory Settings

Feature Key

Bomb Type Select

User Function Setup

Cold Restart


User Factory Settings

User Setup ID

64-1138

Reload Factory Default Settings

Reload User Default Settings

Save User Default Settings





Calibration Data & Controls

Calibration Run Limit

10

EE Max Std Deviation

0.0

Heat of Combustion
of Standard

6318.4

Bomb Service Interval

500

Use Bomb

1


Bomb 1 Through 4

EE Value

940

Protected EE Value

OFF


Thermochemical Corrections

Standardization

Fixed Fuse Correction

ON 50.0

Fixed Acid Correction

ON 8.0

Fixed Sulfur Correction

ON 0.0


Determination
Fixed Fuse Correction

ON 50.0

Fixed Acid Correction

ON 8.0

Fixed Sulfur Correction

OFF 0.0

Calculate Net Heat of Combustion

OFF


Calculation Factors

Acid Value is Nitric Acid Only

ON

Acid Multiplier

0.0709

Sulfur Value is Percent

ON

Sulfur Multiplier

0.6238

Fuse Multiplier

1.0

Use Offset Correction (ISO)

OFF

Offset Value

0.0

Fixed Hydrogen

OFF 0.0

Hydrogen Multiplier

50.68

Dry Calculation

OFF


Dry Calculation
Dry Calculation

OFF

Fixed Moisture

OFF 0.0%

Moisture Multiplier

5.83