beautypg.com

Parr Instrument 6400 User Manual

Page 101

background image

6400

Calorimeter Instruction Manual

14-2

Table 3 - Default Settings


Data Entry Controls

Prompt for Bomb ID

ON

Weight Entry Mode

Touch Screen

Acid Entry Mode

Touch Screen

Hydrogen Entry Mode

Touch Screen

Auto Sample ID Controls

ON

Sample Weight Warning above

2.0

Spike Weight Entry Mode

Touch Screen

Sulfur Entry Mode

Touch Screen

Moisture Entry Mode

Touch Screen

Auto Preweigh Controls

ON


Auto Sample ID Controls

Automatic Sample ID

ON

Automatic Sample ID Number

1

Automatic Sample ID Increment

1


Auto Preweigh Controls

Automatic Preweigh ID

ON

Automatic Preweigh ID increment

1

Automatic Preweigh ID Number

1


Reporting Controls

Report Width

40

Automatic Reporting

ON

Auto Report Destination

Printer

Individual Printed Reports

OFF

Edit Final Reports

OFF

Recalculate Final Reports

OFF

Use New EE Values
in Recalculation

OFF


Communication Controls

Printer Port (RS232)

Balance Port (RS232)

Network Interface

Printer Destination

Local (RS232)

Bar Code Port (RS232)

Network Data Devices



Printer Port Communications

Number of Data Bits

8

Parity None
Number of Stop Bits

1

Handshaking Xon/Xoff
Baud Rate

9600

Printer Type

Parr 1757


Balance Port Communications

Balance Type

Generic

Customize Balance Settings


Balance Port Settings

Number of Data Bits

8

Parity None
Number of Stop Bits

1

Handshaking None
Baud Rate

9600

Data Characters from Balance

8

Data Precision

4

Transfer Time Out

10

Balance Handler Strings

Data Logger

Data Logger

OFF

Interval in Seconds

12

Data Log Destination

Log File and Printer

Data Log Format

Text

Data Log Format

Text