Hanna Instruments HI 4521 User Manual

Page 52

background image

52

Two standard entry types are available: ManualSelection and Custom Standard. The default option is Manual Selection.

A. To calibrate the instrument using

Manual Selection standard entry type:

• Press

. If the instrument was calibrated before and calibration was not cleared, the old calibration can be

cleared by pressing

. After 10 seconds,

will no longer be available.

Note: It is very important to clear calibration history when a new electrode is used

because most errors and warning messages that appear during calibration
depend on calibration history.

• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4 cm

(1½”) into the less concentrated standard solution and stir gently.

• Select the used standard solution concentration with

or

. The “Please wait...” message

will appear on the LCD until the reading is stable or the concentration is validated.

Note: When in ATC mode, if pressing

a pop-up menu will be displayed on the LCD in which the

concentration value can be adjusted using

or

. Press

to save the new

concentration value.

• If the standard solution concentration is validated,

will appear on the LCD. Press

to update

calibration. The calibration point value will be added to the Calibrated Standards list.

• The “Please wait...” message will appear on the LCD for 10 seconds and until the reading is stable, time in

which immerse the ion selective electrode and the temperature probe into the next standard solution and
follow the above procedure or press

to exit calibration.

B. To calibrate the instrument using

Custom Standard entry type:

• Press

. If the instrument was calibrated before and calibration was not cleared, the old

calibration can be cleared by pressing

. After 10 seconds,

will no longer be available.

Note: It is very important to clear calibration history when a new electrode is

used because most errors and warning messages that appear during
calibration depend on calibration history.

• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4

cm (1½”) into the less concentrated standard solution and stir gently.

• When the stability condition is reached and 10 seconds elapsed, a pop-up menu will appear on the LCD

in which the standard concentration value can be set.

• Press

/

to select next/previous entry to be edit. Press

and use

or

to set the desired value. Press

to save the modified value and then press

to store the set concentration value.

65

Load Profile
To load one profile:

• Use

or

to select the

Profile option.

• Press

and use

or

to highlight the

Load Profile option.

• Press

. A list with all customised profiles will be displayed

on the screen.

• Use

or

to select the desired profile and press

to confirm or

to exit without selecting.

Save Current Profile
To save the current profile:

• Use

or

to select the

Profile option.

• Press

and then select

Save Current Profile option. The

Text Editor box will be displayed on the LCD.

• Enter the desired profile name by using

and

to highlight the desired character and then press

to add it to

the text bar. It is also possible to delete the last character by positioning
the cursor on the Backspace character ( ) and pressing

.

• Press

to return to the Profile options.

Note: The saved profile will automatically become the current profile.