Hanna Instruments HI 4521 User Manual

Page 41

background image

41

by pressing

. A pop-up menu will be displayed

on the LCD, in which the slope value can be set using

or

. Press

to save the

modified value or press

to return to the Ion

Constants menu.

Note: If an ISE calibration was performed and a different Ion

Selective Electrode is selected (standard or custom), a
warning message appears on the LCD informing the user to
perform a new calibration or to select the previous ISE in
order to perform accurate measurements.

Concentration Unit
Accessing this option, the user can select the desired concentration unit for the measured ion or chemical compound.
The available concentration units are: ppt (g/L), ppm (mg/L), ppb (µg/L), mg/mL, M (mol/L), mmol/L and %w/v.
To set the Concentration Unit:

• Use

or

to select the Concentration Unit

option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

Sample ID – See

pH Setup section.

Stability Criteria – See

pH Setup section.

Log – See

pH Setup section.

Note: The Logging Data Configuration option includes also the Ion Constants parameter. If you want it to appear

in the log reports, it must be enabled.

Alarm – See

pH Setup section.

Note: The Alarm Limits (Low and High) are set in the selected concentration unit of the measured ion or chemical

compound and are displayed in a scientific format (value and value exponent).

76

Log
This option allows the user to edit the settings related to the logging feature, as following:

Logging Type
Three logging types are available:

Automatic, Manual and AutoHold.

Automatic logging - the readings are logged automatically at constant
time intervals (see

Sampling Period option).

Manual logging ( log on demand)- the readings are logged each time

is pressed.

AutoHold logging - the readings are logged automatically at each
auto-hold event occured.
To set the sample logging type:

• Use

or

to highlight the

Logging Type option.

• Press

and use

or

to choose from

Automatic, Manual or Auto Hold.

• Press

to confirm your selection or press

to

cancel operation.

Logging Data Configuration
This option allows the user to select the parameters that accompany
a logged value:

Date/Time, Calibration Data, Sample ID, Instrument

ID, Operator ID, Company Name, Aditional Info 1 and Aditional Info 2.
To set the logging data configuration:

• Use

or

to highlight the

Logging Data

Configuration option.

• Press

and then use

or

to enable

the parameter by selecting

Yes or to disable it by selecting No for

each option.

• Press

to return to the previous menu.