Three buffer entry types are available: Automatic, Semiautomatic and Manual Selection. The default option is
Manual Selection.

To calibrate the instrument using Manual Selection buffer entry type:

• Press

. If the instrument was calibrated before and calibration was not cleared, the old

calibration can be cleared by pressing

. After 10 seconds,

will no longer be available

from calibration screen.
Note: It is very important to clear calibration history when a new electrode is used

because most errors and warning messages that appear during calibration
depend on calibration history.

• Immerse the pH electrode and the temperature probe approximately 4 cm

(1½”) into a buffer solution of your choice (pH 1.68, 3.00, 4.01, 6.86, 7.01,
9.18, 10.01, 12.45 or a custom buffer) and stir gently. The temperature probe
should be close to the pH electrode.

• Select the used pH calibration buffer using

or

. The “Please

wait...” message will appear on the LCD until the reading is stable or the buffer is validated.

• If the pH buffer is validated,

will appear on the LCD. Press

to update calibration. The

calibration buffer will be added to the Calibrated Buffers list.

• Immerse the pH electrode and the temperature probe into the next buffer solution and follow the above

procedure or press

to exit calibration.

Notes: • The new added calibration point will replace an old one

if the difference between them is ±0.2 pH.

• If the existing stored calibration is full (five calibration

points), a pop-up menu will be displayed on the LCD in
which you can select with

or

keys

the buffer you want to replace with the current buffer.
Press

to delete the selected buffer and then

press

to update calibration with the new

buffer.

• When in MTC mode, if pressing

after entering

pH calibration and while a HANNA pH calibration buffer
with a x.xxx resolution is selected, a pop-up menu will
be displayed on the LCD in which the temperature value can be adjusted using

or

. Press

to save the new temperature value.

73

Set Reminder Period
Daily reminder - the user can set the time of day when the reminder is to appear.
Periodic reminder - the user can set the time from the last calibration (days, hours and minutes) after which the
reminder appears.

To set the reminder period:

• Use

or

to highlight the

Set Reminder Period

option.

• Press

and use

/

to select next /

previous entry to be edited.

• Press

and use

or

to set the desired

value, then press

to save the modified value.

• Press

to return to the previous menu.

Calibration Reminder
This option allows the user to set the calibration reminder as

Daily,

Periodic or Disabled.
To set the calibration reminder:

• Use

or

to highlight the

Calibration re-

minder option.

• Press

to confirm your selection and then use

or

to choose the desired option

.

• Press

to confirm your selection or press

to

cancel operation.