Unrecognized buffer. Please check the buffer or the buffer list (for Semiautomatic and Automatic buffer entry
type): this message appears if the current buffer value is not close to any of the buffers from the buffer list/group.
Check if the current buffer is present in the buffer list or the appropriate buffer group was selected.

The current buffer was already calibrated. Press to recalibrate in this point, or change
the buffer: this message appears when calibration is performed in one of the previous calibrated buffers.
Follow displayed instructions.

71

Compensation Coefficient (only for Linear temperature compensation)
To set the compensation coefficient:

• Use

or

to highlight the

Compensation

Coefficient option.

• Press

and set the desired compensation coefficient

using

or

to increase/decrease the value.

• Press

to save the current value or press

to

cancel operation.

Calibration
Cell Constant
The conductivity probe can be calibrated using the conductivity standards or by entering the cell constant of the
probe by the user.

Using standard solutions:

Using standard solutions:

Using standard solutions:

Using standard solutions:

Using standard solutions:
The meter can be calibrated in a single or multi-points (up to four points), using 6 Hanna standards (84

µS

, 1413

µS

, 5.0

mS

,12.88

mS

, 80.0

mS

, 111.8

mS)

or using the custom standards.

The following options are available for calibration:

Standard Recognition
The user can choose between

Automatic recognition (from 6 Hanna

standards available) or

User Standard (when custom standards are

used for calibration).

• Use

or

to highlight the

Standard Recogni-

tion option.

• Use

or

to choose from

Automatic or User

Standard option.

• Press

to confirm your selection or press

to

cancel operation.