Relative mV

Relative mV

Relative mV

Relative mV

Relative mV MEASUREMENTS

MEASUREMENTS

MEASUREMENTS

MEASUREMENTS

MEASUREMENTS

To measure the Relative mV of a sample:

• Press

and then

to enter

Relative mV

Measure mode.

• The instrument will display the measured Relative mV value

on the LCD, together with a short GLP information about the
last calibration or “Not Calibrated” message if no Rel mV
calibration was performed (no Rel mV offset set).

Notes: • If the measured mV potential is out of range, “-----”

will be displayed on the LCD.

• If pressing

while in

Relative mV Measure

mode, the Relative mV offset screen will be displayed
on the LCD. Press

to clear the existing

Relative mV or use

or

to adjust

the Relative mV value and then press

to save

the current Relative mV. Press

at any time

to return to

Relative mV Measure mode.

• Before performing a Rel mV calibration, make sure

that the electrode is immersed into the sample to be
measured.

67

Temperature
From the

Temperature menu the user can choose the temperature

source and units, as well as the temperature compensation mode,
reference temperature and compensation coefficient.
To access a

Temperature option:

• Use

or

to highlight the

Temperature op-

tion from the

Conductivity Setup menu.

• Press

to access the Temperature option.

Temperature Source
To set the

temperature source:

• Use

or

to highlight the

Temperature Source

option.

• Press

and then use

or

to select

Automatic or Manual temperature source.

• Press

to confirm your selection or press

to

cancel operation.