Isopotential Point
This option allows the user to edit the isopotential point of the electrode used for pH measurements. The isopotential point
is the mV reading for an electrode at which temperature has no effect on the measurement. The ideal electrode has an
isopotential point of 0.0 mV and 7.00 pH, while an actual electrode tipically deviates slightly from the ideal values.
If the actual isopotential pH for an electrode is known, it can be set by accessing this option. The isopotential pH
limits (low and high) will be displayed on the LCD.
Note: If the isopotential point has been modified, recalibration must be performed for the pH channel.
To set the Isopotential Point:

• Use

or

to select the

Isopotential Point option.

• Press

and set the desired isopotential pH value using

or

.

• Press

to save the current value or press

to

cancel operation.

pH Resolution
Accessing this option, the desired pH resolution can be set, with one (x.x), two (x.xx) or three (x.xxx) decimals.
To set the pH Resolution:

• Use

or

to select the

pH Resolution option.

• Press

and use

or

to highlight the

desired option.

• Press

to confirm your selection or press

to

cancel operation.

82

TDS factor

With this option the user can set the TDS factor.

• Use

or

to highlight the TDS options

.

• Press

to confirm your selection and use

or

to increase / decrease the value.

• Press

to confirm your selection or press

to

cancel operation.

Sample ID - see

Conductivity Setup section.

Log - see

Conductivity Setup section.

Alarm - see

Conductivity Setup section.

This manual is related to the following products: