beautypg.com

Source[1..n]:power:state:time – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 456

background image

Referencia de instrucciones SCPI

442

OTDR

Product-Specific Commands—Description

:SOURce[1..n]:POWer:STATe:TIME?

Description

This query returns a value indicating the duration

after which the source will stop emitting light

automatically (auto-off feature).

*RST sets this value to 600 seconds.

Syntax

:SOURce[1..n]:POWer:STATe:TIME?

Parameter(s)

None

Response Syntax

Response(s)

Duration:
The response data syntax for is

defined as a

element.

Duration after which the source will stop

emitting light automatically, in seconds.

Example(s)

SOUR:POW:STAT:TIME 60

SOUR:POW:STAT:TIME? Returns 60

See Also

SOURce[1..n]:POWer:STATe:TIME

SOURce[1..n]:POWer:STATe
SOURce[1..n]:FREQuency:PRF

SOURce[1..n]:FREQuency:PRF:STATe

SOURce[1..n]:FREQuency:BURSt

SOURce[1..n]:FREQuency:BURSt:STATe