beautypg.com

Configure[1..n]:analysis: rbscatter, Configure[1..n]:analysis:rbscatter – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 408

background image

Referencia de instrucciones SCPI

394

OTDR

Product-Specific Commands—Description

:CONFigure[1..n]:ANAlysis:RBScatter

Description

This command sets the Rayleigh backscatter that

will be used for the next acquisition.

*RST reverts this setting to default value.

Syntax

:CONFigure[1..n]:ANAlysis:RBScatter

BS>|MAXimum|MINimum|DEFault

Parameter(s)

RBS:
The program data syntax for is defined

as a element. The

special forms MINimum, MAXimum and DEFault

are accepted on input.

MINimum allows to set the instrument to the

lowest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the parameter.

Sets the Rayleigh backscatter.

Example(s)

CONF:ANA:RBS? Ex.: Returns –79.5

CONF:ANA:RBS –80

CONF:ANA:RBS? Returns –80