Referencia de instrucciones SCPI

378

OTDR

Product-Specific Commands—Description

:CONFigure[1..n]:ACQuisition:MODE

Description

This command specifies the mode that will be

used for the next acquisition.

Acquisition: Allows the OTDR to perform a

standard acquisition.

Auto Setting: Lets the OTDR evaluates the length

of the fiber and finds the appropriate range and

pulse width.

Check First Connector: Used to detect a low

injection level.

Real Time: Used to view sudden changes in the

fiber under test. In this mode, measurements are

not allowed.

*RST sets the current acquisition mode to

ACQUISITION.

Syntax

:CONFigure[1..n]:ACQuisition:MODEACQ

uisition|ASETting|CFConnector|REAltime

Parameter(s)

Mode:
The program data syntax for the first parameter is
defined as a

element. The allowed

DATA> elements for this parameter are:

ACQuisition|ASETting|CFConnector|REAltime.