Referencia de instrucciones SCPI

OTDR

351

Product-Specific Commands—Description

Response(s)

Reflectance:
The response data syntax for is

defined as a

element.

Returns the reflectance value in dB, calculated

using all three markers.

Example(s)

CONF:ACQ:MODE ACQUISITION

INIT

INIT:STAT? Returns 0 when acquisition is

complete.

CALC:REF? TRC1,0,0.1 KM,200 Ex.: Returns –

24.549

CALC:REF? TRC1,0 M,100,200 M Ex.: Returns –

24.549

CALC:REF? TRC1,0 KM,100 M, 0.2 KM Ex.: Returns

–24.549

Notes

See the section on reflectance measurement in

the FTB-7000 Optical Time Domain

Reflectometer user guide.

See Also

CALCulate[1..n]:ANAlysis:[UNIDirectional]

CALCulate[1..n]:EVENt:COUNt?

CALCulate[1..n]:EVENt?

MMEMory[1..n]:LOAD:TRACe

TRACe[1..n]:CATalog?

:CALCulate[1..n]:REFLectance?