beautypg.com

Fetch[1..n]:asetting:range – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 420

background image

Referencia de instrucciones SCPI

406

OTDR

Product-Specific Commands—Description

:FETCh[1..n]:ASETting:RANGe?

Description

This query returns the range found after an

initiate (INIT) command. Note that acquisition

mode (CONF:ACQ:MODE) must be set to

ASETting.

Since *RST clears the range value, the returned

value will be 0.

Syntax

:FETCh[1..n]:ASETting:RANGe?

Parameter(s)

None

Response Syntax

Response(s)

Range:
The response data syntax for is

defined as a

element.

Returns the range, in meters.

Example(s)

CONF:ACQ:MODE ASET

INIT

INIT:STAT? Returns 0 when acquisition is

complete.

FETC:ASET:RANG? Ex.: Returns 1.25E+3

See Also

CONFigure[1..n]:ACQuisition:RANGe?

CONFigure[1..n]:ACQuisition:PULSe?

CONFigure[1..n]:ACQuisition:DURation?

CONFigure[1..n]:ACQuisition:WAVelength:LIST?

CONFigure[1..n]:ACQuisition:RANGe:LIST?

CONFigure[1..n]:ACQuisition:PULSe:LIST?

CONFigure[1..n]:ACQuisition