Plc output to fad-40co input, Plc output to fad-40co input r_dw2, I/o configuration – Flintec FAD-40 User Manual
Page 34
FAD-40 Technical Manual, Rev. 1.06 December 2012
Page 34 of 52
PLC Output to FAD-40CO Input
Dword
D31 D30 D29 D28 D27 D26 D25 D24 D23 D22 D21 D20 D19 D18 D17 D16
D15 D14 D13 D12 D11
D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
RxPDO 1
(R_DW1)
Reserved
RxPDO 2
(R_DW2)
W31
MSB
W30 W29 W28 W27 W26 W25 W24 W23 W22 W21 W20 W19 W18 W17 W16
W15 W14 W13 W12 W11 W10 W9
W8
W7
W6
W5
W4
W3
W2 W01
W00
LSB
RxPDO 3
(R_DW3)
Not used
I/O3
out
I/O2
out
I/O1
out
Not used
Command List
Data Selection
New
CMD
PLC Output to FAD-40CO Input R_DW2
R_DW 2contains
the setting / parameter value (Refer to “Adjust span calibration” and “I/O configuration” in
“Command List”, see PLC Output to FAD-40CO Input R_DW 3).
I/O Configuration
Bit no.
R_DW_2 description
D0…D6
I/O1 output function
0000000
Disable
0000001
Standard setpoint
0000010
Setpoint without hyster.
0000011
Window
0000100
Host output
D7
I/O1 polarity
0
Active Low
1
Active High
D8…D14
I/O2 output function
0000000
Disable
0000001
Standard setpoint
0000010
Setpoint without hyster.
0000011
Window
0000100
Host output
I/O2 input functions
0000101
Zeroing
0000110
Taring
0000111
Clearing tare
0001000
Host input
D15
I/O2 polarity
0
Active Low
1
Active High
D16…D22
I/O3 output function
0000000
Disable
0000001
Standard setpoint
0000010
Setpoint without hyster.
0000011
Window
0000100
Host output
I/O3 input function
0000101
Zeroing
0000110
Taring
0000111
Clearing tare
0001000
Host input
D23
I/O3 polarity
0
Active Low
1
Active High