FAD-40 Technical Manual, Rev. 1.06 December 2012

Page 31 of 52

PLC Output to FAD-40Px Input 3

rd

Dword

Bit no.

3

rd

Dword descriptions

D31…D19 Not in use
D18

Set/reset digital output of I/O3

D17

Set/reset digital output of I/O2

D16

Set/reset digital output of I/O1

D15…D11 Not in use

D10…D6

Command
list

00000 No command is activated
00001 Zero
00010 Tare
00011 Clear
00101 Adjust zero calibration

00110

Adjust span calibration (First load calibration weight on scale then load 2

nd

Dword with span test weight value*, then apply this command with New
CMD)

01000

Operation mode selection (First
load 2

nd

Dword with selected

value, then apply this command
with New CMD)

Load 2

nd

Dword with one of these:

0: Count mode unipolar
1: Count mode bipolar
2: Force mode unipolar
3: Force mode bipolar
4: Weight mode unipolar

01001

mV operation in Count Mode
(First load 2

nd

Dword with selected

value, then apply this command
with New CMD)

Load 2

nd

Dword with one of these:

0: 5 mV
1: 10 mV
2: 15 mV
3: 18 mV

01010

Digital filter
(First load 2

nd

Dword with selected

value, then apply this command
with New CMD)

Load 2

nd

Dword with one of these:

0: Fast

5: Medium


7: Default

9: Slow

01011 I/O Configuration (First, load 2

nd

Dword with desired table)

01100 Setpoint 1 Low for I/O1 (First load 2

nd

Dword with desired value)

01101 Setpoint 1 High for I/O1 (First load 2

nd

Dword with desired value)

01110 Setpoint 2 Low for I/O2 (First load 2

nd

Dword with desired value)

01111 Setpoint 2 High for I/O2 (First load 2

nd

Dword with desired value)

10000 Setpoint 3 Low for I/O3 (First load 2

nd

Dword with desired value)

10001 Setpoint 3 High for I/O3 (First load 2

nd

Dword with desired value)

D5...D1

Read
selected
data

00000 Indicated weight
00001 Gross weight
00010 Tare weight
00011 Indicated weight (Floating point)
00100 Gross weight (Floating point)
00101 Tare weight (Floating weight)
10000 Calibration status

D0

New CMD

Apply commands which are listed in “Command list” (Responds on bit changes)


*Span test weight value has to be entered ignoring the decimal point;
e.g

. “300.00” has to be entered as “30000”