FAD-40 Technical Manual, Rev. 1.06 December 2012

Page 33 of 52

FAD-40CO Output to PLC Input T_DW3

Bit no.

TxPDO 3 (T_DW3) Description

D31…D21

Not in use

D20

Input for I/O2

D19

Input for I/O3

D18

Output for I/O3

D17

Output for I/O2

D16

Output for I/O1

D15…D12

Error codes of
converter

0000

No error found

0001

ADC out

0010

ADC over

0011

ADC under

0100

System error

0101

In setup mode

0110

Low/High voltage error

0111

Instrument not found

D11…D10

Not in use

D9

Operation mode

0

Weight & Force Mode

1

Count Mode

D8

Zero range

0

Weight is out of zero range

1

Weight is in zero range

D7

Indication

0

Gross

1

Net

D6

MD

Motion detection

0

Stable

1

Unstable

D5
D4
D3
D2
D1

Read command
response

00000 Indicated weight
00001 Gross weight
00010 Tare weight
00011 Indicated weight (floating point type)
00100 Gross weight (floating point type)
00101 Tare weight (floating point type)
10000 Calibration Status (Refer to table 8.2)

D0

CMD flag

The command is applied successfully