beautypg.com

Chapter 6 safety device testing, Safety device testing – AERCO BMK 2.0 LN Nat. Gas for Mass. only User Manual

Page 45

background image

SAFETY DEVICE TESTING

CHAPTER 6 SAFETY DEVICE TESTING

6.1 TESTING OF SAFETY DEVICES

Periodic safety device testing is required to
ensure that the control system and safety
devices are operating properly. The Benchmark
2.0LN control system comprehensively monitors
all combustion-related safety devices before,
during and after the start sequence. The
following tests check to ensure that the system
is operating as designed.

Operating controls and safety devices should be
tested on a regular basis or following service or
replacement. All testing must conform to local
codes such as ASME CSD-1.

NOTE

MANUAL and AUTO modes of operation
are required to perform the following
tests. Refer to Chapter 3.

NOTE

It will be necessary to remove the front
door, top panel and side panels from the
unit to perform the following tests.

WARNING

ELECTRICAL VOLTAGES IN THIS
SYSTEM INCLUDE 120 AND 24
VOLTS AC. POWER MUST BE RE-
MOVED PRIOR TO PERFORMING
WIRE REMOVAL OR OTHER TEST
PROCEDURES THAT CAN RESULT
IN ELECTRICAL SHOCK.

6.2 LOW GAS PRESSURE FAULT TEST

Refer to Figure 6-1 and ensure that the ball
valve located at the low gas pressure switch is
closed.

1. Remove the 1/8“ plug from the ball valve at

the low gas pressure switch shown in Figure
6-1.

2. Install a 0 – 16“ W.C. manometer or a W.C.

gauge where the 1/8" plug was removed.

3. Slowly open the ball valve at the low gas

pressure switch.

4. Start the unit in Manual Mode and adjust the

firing rate to between 25 and 30%.

5. While the unit is firing, slowly close the

external manual gas shut-off valve.

6. The unit should shut down and display a

LOW GAS PRESSURE fault message at
approximately 2.6” W.C. The FAULT indi-
cator should also start flashing.

7. Fully open the external manual gas shut-off

valve and press the CLEAR button on the
Control Box.

8. The fault message should clear and the

FAULT indicator should go off. The unit
should restart.

9. Upon test completion, close the ball valve at

the low gas pressure switch and remove the
manometer. Replace the 1/8“ plug removed
in step 1.

Figure 6-1

Low & High Gas Pressure Testing

6.3 HIGH GAS PRESSURE TEST

To simulate a high gas pressure fault, the gas
pressure will be gradually increased using the
adjustment screw on the SSOV regulator. This
adjustment is located behind the brass hex head
cap shown in Figure 4-3. Perform the test as
follows:

1. Refer to Figure 6-1 and remove the 1/8“ plug

from the ball valve at the high gas pressure
switch.

2. Install a 0 – 16” W.C. manometer or W.C.

gauge where the 1/8” plug was removed.

6-1