beautypg.com

Setting, Auto zero/output adjustment, Output adjustment – KEYENCE LB-1000(W) User Manual

Page 10

background image

10

SETTING

1. Adjust the distance between the sensor head and target and confirm

the position with the LASER ON alarm LED. Set the sensor head so
that the laser emitting surface is parallel to the measuring surface of
the target, then secure it in place. (See figure below.)

When the LED lights green, the target is in the center area of the
measuring range.

LB-041: approx. 40 mm
LB-081: approx. 80 mm
LB-301: approx. 300 mm

2. Adjust orientation of the sensor head (for a moving target).

Be sure to mount the sensor head as shown in the right-hand figure
below to ensure stable measurement.

Example: Step detection

Example: Eccentricity measurement

AUTO ZERO/OUTPUT ADJUSTMENT

Press the AUTO ZERO key to reset the output voltage to zero at the
mounting position.

Example: LB-041

AUTO ZERO setting is possible from external terminals (connecting the
6 terminal to a GND terminal).
* The AUTO ZERO function can be used when the DIP switch No. 3 of

the controller is set to FREE (top setting).

Output adjustment

If the output voltage value is not in proportion with the target displace-
ment, adjust the output voltage using the SPAN adjustment trimmer.
Example: LB-041

When a sensor is positioned on both sides of a
target for measurement, be sure to adjust the span
of each sensor separately.

Incorrect

Correct

Incorrect

Correct

V

The LED light green.

Approx.
40 mm

ZERO

Panel meter

*12 mA for

current output.

V

V

40 mm

38 mm

V

SPN-ADJ

adjustment +0.010.

Change by 2 mm

(To change
-2.010 to -2.000.)

Span adjustment

Note

Measuring range

LASER ON
alarm LED

Red

Yellow

Green

Yellow

Red

Center

Distance between sensor head and target

Out of
measuring
range
(Operating range)

Out of
measuring
range
(Operating range)