Holtgreven GSE-375 User Manual
Page 74

68
DISPLAY READS Enter ~ Load ~ xx.xx
11. Enter 100 with the [PRINT] and [UNITS] keys.
DISPLAY READS 100
12. Press [TARE] to establish span.
DISPLAY READS 100.00
13. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 100.00
14. Press [TARE] to accept calibration.
DISPLAY READS Enter ~ =Stor
15. Press [TARE] to save calibration.
DISPLAY READS Enter ~ =End
16. Press [TARE] to exit calibration.
DISPLAY READS 100.00
17. Remove the calibration weight.
DISPLAY READS 0.00
Reset Calibration Gain Factors Example (375):
1. Press
100 [SELECT].
DISPLAY READS Setup
2. Press
54321 [SAMPLE/Enter].
DISPLAY READS Fast ~ Cal~First ~ Zero? ~ xx.xx
3. Remove any load on the scale.
DISPLAY READS First ~ Zero? ~ xx.xx
4. Press
[SAMPLE/Enter] to establish zero.
DISPLAY READS Enter ~ Load? ~ 0.00
5. Place a 100lb test weight on scale.
DISPLAY READS Code03
6. Press
[ZERO].
DISPLAY READS First ~ Zero? ~ -0.26
7. Remove any load on the scale.
DISPLAY READS First ~ Zero? ~ -0.42
8. Press
[SAMPLE/Enter] to establish zero.
DISPLAY READS 0.00
9. Pause for motion delay.
DISPLAY READS Enter ~ Load ~ 0.00
10. Place a 100lb test weight on scale.
DISPLAY READS Enter ~ Load ~ xx.xx
11. Enter 100.
DISPLAY READS 100
12. Press [SAMPLE/Enter] to establish span.
DISPLAY READS 100.00
13. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 100.00
14. Press [SAMPLE/Enter] to accept calibration.
DISPLAY READS Enter ~ =Stor