7. Press
[TARE] to use last zero.
DISPLAY READS Enter ~ Load? ~ 9970
8. Enter
10000.
DISPLAY READS 10000
9. Press
[TARE] to establish span.
DISPLAY READS 10000.
10. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 10000.
11. Press [TARE] to accept calibration.
DISPLAY READS Enter ~ =Stor
12. Press [TARE] to save calibration.
DISPLAY READS Enter ~ =End
13. Press [TARE] to exit calibration.
DISPLAY READS 10000.
14. Remove the calibration weight.
DISPLAY READS 0.
Last Zero Calibration With Weight Already Applied Example (375):
1. Remove any load on the scale.
DISPLAY READS 10.
2. Press
[ZERO] to zero the scale.
DISPLAY READS 00.
3. Apply a 10000 lb test weight to verify calibration.
DISPLAY READS 9970.
4. Press
100 [SELECT].
DISPLAY READS Setup
5. Press
54321 [SAMPLE/Enter].
DISPLAY READS Fast ~ Cal~First ~ Zero? ~ 9930.
6. Press
[SELECT].
DISPLAY READS Last ~ Zero? ~9930.
7. Press
[SAMPLE/Enter] to use last zero.
DISPLAY READS Enter ~ Load? ~ 9970.
8. Enter
10000.
DISPLAY READS 10000
9. Press
[SAMPLE/Enter] to establish span.
DISPLAY READS 10000.
10. Pause for motion delay.
DISPLAY READS Cal ~ Good? ~ 10000.
11. Press [SAMPLE/Enter] to accept calibration.
DISPLAY READS Enter ~ =Stor
12. Press [SAMPLE/Enter] to save calibration.
DISPLAY READS Enter ~ =End
13. Press [SAMPLE/Enter] to exit calibration.
DISPLAY READS 10000.
14. Remove the calibration weight.
DISPLAY READS 00.