Holtgreven GSE-375 User Manual
Page 32
![background image](/manuals/657671/32/background.png)
26
P233
Transmit Selection (Comm 2) (Selection)
Select desired print output (1 - 16 or custom). The transmission will be initiated by the selected
print operation (P210) and / or the Remote Key selection (P800). See page 29 for details on
preset formats or page
31 for details on custom transmit.
P250 RS-485 Multi-Drop Network (Toggle)
Enable / disable the RS-485 multi-drop network option. Requires that an RS-485 option board be
installed. This option allows up to 250 RS-485 devices to be networked together in either a half
duplex or full duplex wiring scheme. See the RS-485 Multi-Drop Network Setup and Operation
section beginning on page
35 for complete details on RS-485 setup and operation.
P251
Address (Key in) (Only displayed if P250 is enabled)
Specifies the address of the Model 370/375 for RS-485 multi-drop communications. Allowed
choices are 0 (disabled) and 4 - 254.
P260
Master, Remote Display or Safe Area Hub (Selection)
Select the mode of operation.
P290
Echo Display (Toggle)
Determines which comm port will be echoed to another device. Selections are None, port 1, port
2.
P291
Start Character (Key-in)
Set the start character for the master indicator. Choices are 0 – 255. Default value is 2 (STX).
The choice must match the start character at P274 of the remote display (slave). A value of 0
means that a byte with the value 0 will be sent for the character. Avoid using values of 17 or 19
when using software flow control. The values of 17 and 19 are Xon and Xoff codes and could
cause conflict.
P292
End Character (Key-in)
Set the end character for the master indicator. Choices are 0 – 255. Default value is 3 (ETX). The
choice must match the end character at P275 of the remote display (slave). A value of 0 means
that a byte with the value 0 will be sent for the character. Avoid using values of 17 or 19 when
using software flow control. The values of 17 and 19 are Xon and Xoff codes and could cause
conflict.
P296
Echo Style (Toggle)
Set the style of indicator the display will be echoed to. Choose from 300 or 60. The 300 choice is
for 350, 355, 350 I.S., 355 I.S., 351, 370, 375, ABS 4100. The 60 choice is for 460 series, 560
series and 660 series indicators.
P410 Euro
(Toggle)
Enable/disable OIML legal-for-trade restrictions. To enable this parameter key in 9991
[SAMPLE/Enter]. To disable key in 9990 [SAMPLE/Enter].
If P410 is enabled the following will apply:
• An over-load condition will result when the gross weight exceeds nine graduations over
the full scale capacity.
Choice Number
Selection Name
Description
P260.00
350
Sets indicator as the master device. Model 350/355/370/375.
P260.01
R-dsp
Sets the indicator as a slave device. P261 – P283 will be
available.
P260.02
Hub
Sets the indicator to a safe area hub.
P260.03
Cntrl
Sets the indicator to the master controller for fiber-optic
interface to the safe area hub.