2.8.3 Peak Capture Rate: 80 Hz
2.8.4 Input range: -15mV to +15mV
2.8.5 Output code: 1mV input between S+ and S- of load cell connector will output about 100,000 raw
Counts.
2.8.6 With Hardware low pass filter and 2 programmable digital low pass filters
2.9 Capacity and Division: Programmable
2.9.1 Max display range: -999,999 to 999,999
2.9.2 Max number range of all units = 100-100,000
2.9.3 Recommended Display Sensitivity: >0.5uV/ display division
2.10 Accuracy: ≤0.01%
2.11 Calibration Method:
2.11.1 Software calibration with long-term storage in EEPROM
2.11.2 Provides smooth curve fit through linearization points.
2.11.3 Calibration can be done under kgf or lbf weight unit with 10% -100%FS standard weight
2.12 Real Clock: built-in nonvolatile real time & date
2.13 Remote Digital Input:
2.13.1 There are four external inputs
2.13.2 Function: each input can be defined to HOLD, ZERO, TARE, PRNT, UNIT, OFF or NONE
2.14 Digital Output (Set point Output):
2.14.1 There are two configurable outputs
2.14.2 Each digital output combined with a digital comparator: if current force is over or less than the set
force point, the corresponding output will change its output electronic state according to the
pre-configuration.
2.15 Analog Voltage Output:
2.15.1 The output voltage is 0-2.5V, the output current should be limited 5mA by external circuit.
2.15.2 The accuracy of this analog voltage is 0.01% after calibration
2.16 Other Main Function:
2.16.1 Programmable Zero Range
2.16.2 Programmable automatic zero point tracking
2.16.3 Programmable motion detection window
2.16.4 Programmable auto-power off time, adjustable LED brightness
2.16.5 Available Measure Unit: kgf, lbf, Newton
2.16.6 Programmable serial output content