Dillon FI-521 User Manual
Page 15
FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
15
5.3 USER Submenu:
To enter this parameter, press and hold the SELECT key for 3 seconds until display shows CONFIG. Press
CELL key to USER. Press ZERO key to RESET and press CELL key to COM1.
USER
Sub-
Menu1
SubMenu2 Option Default
Remark
RESET NO
NO
reset user parameters to default setting
YES
COM1 BAUD.RT 1200
9600
selection of com1's baud rate
2400
4800
9600
19200
38400
57600
BYT.FMT
8N1
8N1
selection of com2's byte format:
(1) 8N1=8 data bits, No parity check bit, 1 stop bit;
(2) 7O1=7 data bits,1Odd parity check bit, 1 stop
bit;
(3)7E1=7 data bits, 1 Even parity check bit, 1 stop
bit;
7O1
7E1
OUT.MOD
NONE
PRT.CMD
Selection com1 output mode:
(1)NONE= COM1 disabled;
(2)CONT=continuously output;
(3)PRINT=output after PRINT key pressed;
(4)CMD=output after a request command is
received;
(5)PRT.CMD= output after PRINT key pressed or
request command received;
(6)STABLE=output after indicator is stable;
Note: use PRINT or CMD to output data, the
indicator must be stable.
CONT
CMD
PRT.CMD
STABLE
LAYOUT
MULTPL
MULTPL
com1 output content and format set:
(1)MULTPL= the following selected item in OUT1
will be output use defined format;
(2)SINGLE= only displayed content and current
status will be output, it’s compatible with
NCI-SCP01
SINGLE
OUT1 TITLE YES
NO
Yes/No=enable/disable output prompt of every
output line
NO
IND.ID
YES
NO
Yes/No=enable/disable output indicator's ID
number, Prompt is “IND. ID”
NO
CELL.No
YES
NO
Yes/No=enable/disable output load cell’s ID
number, Prompt is “CELL No.”
NO
CELL.TYP
YES
NO
Yes/No=enable/disable output load cell’s type,
Prompt is “CELL TYPE”
NO
DSP.POL
YES
NO
Yes/No=enable/disable output load cell’s measure
type, Prompt is “DSP.POL”
NO
FORCE
YES
YES
Yes/No=enable/disable output gross force.
Prompt is “FORCE”
NO
PEK.VLY
YES
NO
YES/NO=enable/disable output current captured
peak value and valley value; prompt is “PEAK”
and “VALLEY”
NO