5. Operation Menu Structure

5.1 Main menu:

(Cell)

(Cell) (Cell) (Cell) (Cell)

NOTE:

(1) Each LOAD CELL has its own CONFIG, CAL, IN.OUT parameters, so, before you enter

this main menu select, which load cell will be configured with the CELL key!

(2) The parameters of USER, MISC, and TEST are the same for all load cells

(3) When “Lo.VoL” or “Lo.BAT” is displayed (the voltage to PCB is low), CONFIG, USER,

CAL, IN.OUT menu can not be entered and edited.


5.2 CONFIG Submenu:
5.2.1 Access the set up mode by pressing and holding the SELECT key for 3 seconds, until display shows

CONFIG.

5.2.2 Press

the

ZERO key once (ZERO key is your Enter key)

5.2.3 Display will show CFG.ON or CFG.OFF

a. If it shows CFG.OFF, then you will need to move the switch in the back of the indicator,

through the CAL. Switch cover. Cycle power of the indicator and start at step #1 again.

5.2.4 Press

the

ZERO key once. Display shows RESET

5.2.5 Pres

the

CELL key until display shows CELL.TY

5.2.6 Press

ZERO key once. Display shows TS.CP

5.2.7 Press

ZERO key once. Display shows CELL.TY

5.2.8 Press

CELL key once. Display shows DSP.POL

5.2.9 Press

ZERO key once. (Select NEG or POS)

5.2.10 Press ZERO to accept
5.2.11 Press CELL key to PRIM.N
5.2.12 Press ZERO key. Display shows 10,000

a. This is number of Divisions. Use the chart below as a reference

CAPACITY is based on RESOLUTION

Capacity & Count by

Required (LBS)

PRIM.N PRIM.D

500 X 0.1

5,000

.1

2,000 X 0.5

4,000

.5

5,000 X 1

10,000

.5

10,000, X 2

10,000

1

50,000 X 10

5,000

10


** Take your Capacity and divide by your desired count-by (PRIM.D) will give you your PRIM.N number
** Example = 5,000 lb load cell divided by .5 lbs will give you 10,000
5.2.13 Press UNIT key to change to the desired number you want to change. Stop when your digit is flashing.
5.2.14 Press CELL key or PRINT key to change a number
5.2.15 Once you have set your DIVISION press ZERO to accept. Display shows PRIN.N

CONFIG

USER

CAL

IN.OUT

MISC

TEST