FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

25

a. If yes is selected, the calibration will be saved and exit to the force mode when the load is

removed.

b. If no is selected, it will go to the next step.

7.15 Press the ZERO key to confirm

7.16 The first default standard weight is displayed after Cal.P3being shown. The default standard weight is

100 % of full scale. Use the SELECT, PRINT, CELL, UNIT keys to input the value of the loaded weight.

(more than 10% of full scale weight, and larger than the weight used on CAL.P2) onto scale

7.17 Press the ZERO key to confirm
7.18 When calibration is complete CAL.ENDwill be displayed. The indicator will restart, remove the load

from the scale and the indicator will go back to the normal force mode.

8. Force Fine-tune

With this function, the user can adjust displayed force a little, and no need standard weight. But please
note:

(1)The indicator must have been calibrated before this
(2)The range of adjustment is “(current displayed force) x (0.9-1.1)”. it means the range is about ±10%
(3)The “CONFIG-F.ADJ=YES” must be set.
(4) The current unit should be the unit in PRIM.U, and the tare weight is 0.

8.1 To enter this mode, turn on indicator, after indicator displaying 0 force, put a load (suppose: its correct force

is 1230.0lb) onto load cell, then indicator will display the load’s force, say “1234.5 lb”. Press down
ON/OFF and ZERO at same time until first digit flashes, this means indicator has entered into “force
fine-tune” mode.

8.2 Use to input correct force (1230.0). Press the ZERO key, the active correct force will be

displayed and Display will stop flashing. After this, displayed force will be adjusted by this ratio
(1230.0/1234.5) and this ratio will be active until next modification on it.

8.3 To remove the effect of this ratio, there’re two ways: “8.3.1 way” and “8.3.2 way”

8.3.1 Do standard calibration, like in “6.CALIBRATION”
8.3.2 Remove all weight on scale, Press ZERO to make 0 displayed, Put a load onto load cell, a number

will be displayed, suppose it’s 1230.0lb ; Press down ON/OFF and ZERO at same time until first
digit flashes, this means indicator has entered into “force fine-tune” mode.

Press key, the displayed weight will be restored to 1234.5, and then press key to

confirm and exit to normal weighing mode.


9. View ADC output Code
9.1 In this mode, you can examine the stability of weighing system, the increment value of ADC output code

corresponding to the loaded weight.

Note:

The increment of ADC code for FS weight must be larger or equal to 10 times of selected

display division; otherwise, the calibration cannot be properly completed. e.g. The display

division is 0.1kg. Load 100kg standard weight on the platform, the increment of ADC code

should be at least more than 10x100kg/0.1kg= 10x1000=10000. In this case, the indicator can

be calibrated. Otherwise, smaller division needs to be Chosen.

The variation of ADC code should be small; otherwise, the calibration cannot properly complete

also.