FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

18

OTHER

NLD.RNG 1-255

10

1-255=the range of force is ±1-255d; when
current force is less than this value, the load cell
can be regarded as no load on it. It must be bigger
than (CONFIG.MOTION).

CMD.SRC

NONE

COM1

Source of the executed command selection:
NONE=no command will be executed;
COM1= command from COM1 will be executed;
COM2= command from COM2 will be executed;
COM1-2= command from COM1 or COM2 will be
executed;

COM1

COM2

COM1-2

A.OFF.T 0-255

0

Auto off time:
0=not auto power off;
1-255=auto power off after 1-255 minutes, in this
period, it should be no operation or no weight
changing

OFF.MOD

OFF

OFF

Auto off mode:
(1)OFF=turn off instrument;
(2)DSP.TIM= display time;
(3)DSP.DAT=display date;
(4)AC.TIME=turn off when only battery is used,
display time when AC adaptor is used;
(5)AC.DATE=turn off when only battery is used,
display date when AC adaptor is used;

DSP.TIM

DSP.DAT

AC.TIME

AC.DATE

LED.BRT

BRT1-2-3-4-

5

BRT4

LED brightness level set:
BRT1-BRT2-BRT3-BRT4-BRT5=low - middle -
high; only available on LED version

IND.ID

000000 -

999999

123456

Indicator’s ID number: 000000-999999