FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
18
OTHER
NLD.RNG 1-255
10
1-255=the range of force is ±1-255d; when
current force is less than this value, the load cell
can be regarded as no load on it. It must be bigger
than (CONFIG.MOTION).
CMD.SRC
NONE
COM1
Source of the executed command selection:
NONE=no command will be executed;
COM1= command from COM1 will be executed;
COM2= command from COM2 will be executed;
COM1-2= command from COM1 or COM2 will be
executed;
COM1
COM2
COM1-2
A.OFF.T 0-255
0
Auto off time:
0=not auto power off;
1-255=auto power off after 1-255 minutes, in this
period, it should be no operation or no weight
changing
OFF.MOD
OFF
OFF
Auto off mode:
(1)OFF=turn off instrument;
(2)DSP.TIM= display time;
(3)DSP.DAT=display date;
(4)AC.TIME=turn off when only battery is used,
display time when AC adaptor is used;
(5)AC.DATE=turn off when only battery is used,
display date when AC adaptor is used;
DSP.TIM
DSP.DAT
AC.TIME
AC.DATE
LED.BRT
BRT1-2-3-4-
5
BRT4
LED brightness level set:
BRT1-BRT2-BRT3-BRT4-BRT5=low - middle -
high; only available on LED version
IND.ID
000000 -
999999
123456
Indicator’s ID number: 000000-999999