FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

12

5.2.16 Press CELL key. Display shows PRIM.D
5.2.17 Press ZERO key once. Display shows 1
5.2.18 Press CELL key or PRINT key to change the digit

a. Once you make a choice press ZERO key to accept

5.2.19 Press CELL key. Display shows PRIN.
5.2.20 Press ZERO key. Display shows LbF

a. Press

CELL key to toggle between kgF or LbF

5.2.21 Press ZERO key to accept
5.2.22 Press ON/OFF key twice
5.2.23 Display will cycle software version, then capacity, then go to normal weighing mode.
5.2.24 Move CAL. Switch back to CFG.OFF (if you want to lock calibration)

CONFIG

SubMenu1 SubMenu2

Option

Default

Remark

CFG.ON

seal switch is on

CFG.OFF

Seal switch is off

RESET

NO

NO

reset configure parameters to default setting

YES

CELL.TY

Tension

Tension

Select the connected load cell is used for
tension or compression force

Compres

TS.CP

DSP.POL

Pos

Pos

Select the display polarity sign of positive or
negative for the connected load cell. When
CELL.TY = TS.CP, “DSP.POL=POS” indicates
that the inner code increased direction is
tension direction; “DSP.POL=Neg” indicates
that inner code decreased direction is tension
direction.

Neg

PRIM.N

100 – 100,000

10,000

the division number under primary unit

PRIM.D

0.0001,
0.0002,
0.0005,

0.001,
0.002,
0.005,

0.01,
0.02,
0.05,

0.1, 0.2, 0.5,

1, 2, 5,

10, 20, 50

1

The division value under primary unit; the
division value under second unit is
automatically determined by indicator according
to the division value under primary unit.

PRIM.Ut

KGF

LBF

Select the primary unit from kg or lb, the second
unit is the lb if kg selected as primary unit or kg
if lb selected as primary unit; the calibration
standard weight must be in primary unit!

LBF