beautypg.com

Dillon FI-521 User Manual

Page 14

background image

FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

14

FILTER

FLT1.TH 0-255

40

Enter digital filter1 threshold:
0=no filter1;
1-254=filter1 be used only when vibration in
±0.5d*(1-254) ;
255= filter1 be always used

FLT1.ST 1-64

8

Digital filter1 intensity: 1-64 ADC’s data will be
averaged

FLT2.TH 0-255

20

Enter digital filter2 threshold: 0=no filter2;
1-254=filter2 be used only when vibration in
±0.5d*(1-254) ; 255= filter2 be always used

FLT2.ST

0-255

160

Digital filter2 intensity: 0-255=weak to strong

FUNC F.ADJ YES

NO

Enable or disable fine-adjust force number in
normal force measure mode

NO

PEAK Threshd

5-65535 10

5-65535d, When reverse varying(Threshold) of
force is over this value, one peak is occurred
and will be captured