FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
14
FILTER
FLT1.TH 0-255
40
Enter digital filter1 threshold:
0=no filter1;
1-254=filter1 be used only when vibration in
±0.5d*(1-254) ;
255= filter1 be always used
FLT1.ST 1-64
8
Digital filter1 intensity: 1-64 ADC’s data will be
averaged
FLT2.TH 0-255
20
Enter digital filter2 threshold: 0=no filter2;
1-254=filter2 be used only when vibration in
±0.5d*(1-254) ; 255= filter2 be always used
FLT2.ST
0-255
160
Digital filter2 intensity: 0-255=weak to strong
FUNC F.ADJ YES
NO
Enable or disable fine-adjust force number in
normal force measure mode
NO
PEAK Threshd
5-65535 10
5-65535d, When reverse varying(Threshold) of
force is over this value, one peak is occurred
and will be captured