FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

13

SECOND.N

100-125,000

10,000

the division number under second unit ,the max
is 1.25*(PRIM.N),

NEWTON.N

100 – 125,000

10,000

the division number under the NEWTON
unit ,the max is 1.25*(PRIM.N),

UNITS

0-6 6

units that can use UNIT key to select:
0=kgf, 1=lbf, 2=N(Newton), 3= kgf,lbf,
4=kgf,N, 5=lbf,N, 6=kgf,lbf,N
refer to section5.12 for some limitation

MOTION

1-255

4

Check motion window: 1-255=±0.25d *(1-255)

OVER.LD

0-100 0

over load display limitation:
0=FS+9d,
1-100=101%FS to 200%FS

ABUSE 0-255 20%

FS+(0%FS to 255%FS), If the force has exceed
this level, it’s considered to be abusive to the
load cell

ZRO.PNT

P.IZSM 0-100 10

Initial zero(power on zero) point range:
0=no limitation,
1-100= (calibration zero point) + 1%FS to
(calibration zero point) +100%FS

N.IZSM 0-100 10

Initial zero(power on zero) point range:
0=no limitation,
1-100= (calibration zero point) - 1%FS to
(calibration zero point) -100%FS

IN.IZSM

FORCE

FORCE

Which force will works as initial zero point when
force is in IZSM range:
FORCE= current force ;
CAL.ZRO= calibration zero;
LAST.ZRO=switch-off zero force

CAL.ZRO

LAST.ZRO

OV.IZSM

DSP.OVR

DSP.OVR

Choose which force will works as initial zero
point when force is over IZSM range:
DSP.OVR=display initial zero point is over;
FORCE= current force;
CAL.ZRO= calibration zero;
LAST.ZRO=switch-off zero force

FORCE

CAL.ZRO

LAST.ZRO

SAZSM 0-100

2

Zero key range:
0=no limitation,
1-100= (initial zero point) ±1%FS to (initial
zero point) ±100%FS

AZSM 0-255 2

Zero tracking window:
0=0d, no tracking;
1-255=±0.25d*(1-255)