Dillon FI-521 User Manual
Page 28
FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
28
20.1 Before test the receiving function of COM1 or COM2, a cable is need to connect a PC and this indicator,
and a software that is similar with Super Terminal of Windows is also need to run on PC to send bytes to
this indicator. Please note: testing uses USER-COM1/2-BAUD.RT setting baud rate and 8N1 byte format,
Hex data.
20.2 Press down SELECT more than 3s to enter MAIN MENU mode, using PRINT or CELL key to select
TEST-COM1.RD or TEST-COM2.RD item, press ZERO to enter test COM1/2 receiving function, and
rd1.-- or rd2.--will be displayed first.
20.3 In this mode, received hex data (0x00 – 0xff) will be displayed on -- position.
20.4 Press SELECT key to return to last menu item, press ON/OFF key to prepare to exit this mode
21. Serial Port1/2(COM1/2) Transmitting Test
21.1 Before test the transmitting function of COM1 or COM2, a cable is need to connect a PC and this
instrument, and a software that is similar with Hyper Terminal of Windows is also need to run on PC to
receive bytes from this instrument. Please note: testing uses USER-COM1/2-BAUD.RT setting baud rate
and 8N1 byte format, Hex data.
21.2 Press down SELECT more than 3s to enter SELECT mode, using PRINT or CELL key to select
TEST-COM1.TD or TEST-COM2.TD item, press ZERO to enter test COM1/2 transmitting function, and
td1.-- or td2.-- will be displayed first.
21.3 In this mode, transmitted hex data (0x00 – 0xff) will be displayed on -- position, and PRINT, CELL, UNIT,
ZERO keys can be used to modify transmitted data.
21.4 Press SELECT key to return to last menu item, press ON/OFF key to prepare to exit this mode
22. Remote Input Function Selection
22.1 Press down SELECT until CONFIG is shown, using PRINT or CELL key to select IN.OUT-INPUT item,
press ZERO to display original function of input line1.It is displayed in “Inx.yyyy) format. X(=1,2,3,4) is the
number of input line, yyyy(=NONE, SELE, CELL, PRNT, UNIT, ZERO, OFF) is the function of input line x.
Every input function can be defined to SELECT,CELL,PRINT,UNIT,ZERO, or OFF.
22.2 Use CELL key to change input line; Use PRINT key to select the function of this line, Use ZERO key to
confirm and save.
22.3 Press SELECT key to return to last menu item, press ON/OFF key to prepare to exit this mode
23. Output Setting
23.1 Press down SELECT until CONFIG is shown, using PRINT or CELL key to select IN.OUT-OUTPUT item,
press ZERO to enter setting mode.
23.2 Use CELL or PRINT key to select MODE (setting output mode) or SPF1 or SPF2 (“Set Point Force”,
setting the corresponding data of output). Use ZERO to confirm.
23.3 When “OUTx.yz” is shown, Use CELL key to change output line number(x), Use PRINT key to change
level after power on (y) and after an event trigger(z); x=1-2, y=0-1, z=0-1; Normally, y/z=0 will make
external relay closed, and y/z=1 make the relay open. y=0/1 means after the indicator power on, the
output level is 0/1, z=0/1 means if current display force is over the setting data (SPF1/2), the output level is
0/1. Use ZERO to confirm.