26.3.11 AVOUT Connect on Analog Voltage Output Sub-Board:

CONNECTED PINS

FUNCTION

1

Analog Voltage Output +

2

Analog Voltage Output – (Internal Power GND)

Note:

The current draw from 1-2 connector should be less than 2.5mA, that means the resistor added
between 1-2 connector should more than 1K

M

!





26.3.12 XIN (Remote Input) Connector:

26.3.12.1 Pin Definition

PIN #

DEFINITION

IN/OUT/POWER

ELECTRICAL LEVEL

1

Internal Power output +

Power output

5 - 8Vdc (≤0.1A)

2 Input1+

input

5-24Vdc

3 Input1-

input

0Vdc

4 Input2+

input

5-24Vdc

5 Input2-

input

0Vdc

6 Input3+

input

5-24Vdc

7 Input3-

input

0Vdc

8 Input4+

input

5-24Vdc

9 Input4-

input

0Vdc

10

Internal Power output -

Power GND

0 Vdc




26.3.12.2 Connection Example When External Power Supply:


The remote keys RK1-RK4 also can be transistor, and the remote keys connection is isolated to internal circuit.