FI-521 Indicator Operation & Service Manual

────────────────────────────────────────────────────────

20

5.5 IN.OUT Submenu:

IN.OUT

SUB-

MENU1

SUB-

MENU2

SUB-

MENU3

DEFAULT REMARK

INPUT INx.

yyyy

IN1.SELE
IN2.CELL
IN3.UNIT
IN4.ZEO

Select function(yyy) of each remote input(x)
X=1-4; yyyy=SELE(select)/ CELL(cell)/ PRNT(print)/
UNIT(unit)/ ZERO(zero)/ OFF/ NONE(none); normally,
one remote input is low (0), that means one external
switch is closed

OUTPUT

MODE OUTx.yz

OUT1.00
OUT2.00

Select output(x) level after power on(y) and after an event
trigger(z); x=1-2, y=0-1, z=0-1; Normally, y/z=0 will make
external relay closed, and y/z=1 make the relay open

SPF1

0kgf

Select the force unit of OUTPUT1 referred:
kgf→lbf→N→kgf; and Input the force number of
OUTPUT1 referred: 0 to FS

SPF2

0kgf

Select the force unit of OUTPUT2 referred:
kgf→lbf→N→kgf; and
Input the force number of OUTPUT2 referred: 0 to FS

AVOUT

MODE

OFF

OFF

OFF/ON=analog voltage output is off/on

ON

2.5V-F

3000kgf

Select the controlling force unit: kgf→lbf→N→kgf
Input the force value or capacity that you want to the
2.5VDC reference: 10%FS to 120%FS. Default is 10,000.

CAL.VO

0.0000 V

------------

To set and check the output voltage of
0.0000/1.2500/2.5000VDC

1.2500 V

2.5000 V

4.6 MISC Submenu:

MISC

SUBMENU1 REMARK

CODE

display ADC's code, this code can be after no-filter, filter1 or filter2; or display input
signal in mV/V; details refer to section8

VOL

display voltage; calibrate voltage; set full charged voltage and low battery voltage;
details refer to section9