Dillon FI-521 User Manual
Page 20
FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
20
5.5 IN.OUT Submenu:
IN.OUT
SUB-
MENU1
SUB-
MENU2
SUB-
MENU3
DEFAULT REMARK
INPUT INx.
yyyy
IN1.SELE
IN2.CELL
IN3.UNIT
IN4.ZEO
Select function(yyy) of each remote input(x)
X=1-4; yyyy=SELE(select)/ CELL(cell)/ PRNT(print)/
UNIT(unit)/ ZERO(zero)/ OFF/ NONE(none); normally,
one remote input is low (0), that means one external
switch is closed
OUTPUT
MODE OUTx.yz
OUT1.00
OUT2.00
Select output(x) level after power on(y) and after an event
trigger(z); x=1-2, y=0-1, z=0-1; Normally, y/z=0 will make
external relay closed, and y/z=1 make the relay open
SPF1
0kgf
Select the force unit of OUTPUT1 referred:
kgf→lbf→N→kgf; and Input the force number of
OUTPUT1 referred: 0 to FS
SPF2
0kgf
Select the force unit of OUTPUT2 referred:
kgf→lbf→N→kgf; and
Input the force number of OUTPUT2 referred: 0 to FS
AVOUT
MODE
OFF
OFF
OFF/ON=analog voltage output is off/on
ON
2.5V-F
3000kgf
Select the controlling force unit: kgf→lbf→N→kgf
Input the force value or capacity that you want to the
2.5VDC reference: 10%FS to 120%FS. Default is 10,000.
CAL.VO
0.0000 V
------------
To set and check the output voltage of
0.0000/1.2500/2.5000VDC
1.2500 V
2.5000 V
4.6 MISC Submenu:
MISC
SUBMENU1 REMARK
CODE
display ADC's code, this code can be after no-filter, filter1 or filter2; or display input
signal in mV/V; details refer to section8
VOL
display voltage; calibrate voltage; set full charged voltage and low battery voltage;
details refer to section9