Dillon FI-521 User Manual
Page 19
FI-521 Indicator Operation & Service Manual
────────────────────────────────────────────────────────
19
5.4 CAL Submenu:
To enter this parameter, press and hold the SELECT key for 3 seconds until display shows CONFIG. Press
CELL key (2x) to CAL. Press ZERO key to enter Cal Menu.
CAL
SUBMENU1 SUBMENU2
OPTION
REMARK
CAL.ON
Seal switch is on
CAL.OFF
Seal switch is off
ZERO
Only do zero point calibration, then go to CAL.END to
end. (Only needed when a zero shift has occurred.)
LINE CAL.P0
Line calibration point0: do zero point calibration,
this point can't be omitted.
CAL.P1
Line calibration point1: do first weight point
calibration, this point can't be omitted and standard
weight must be over 10%FS.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P2
Line calibration point2: do second weight point
calibration, standard weight must be over 10%FS and
be larger than it in CAL.P1, this point can be omitted.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P3
Line calibration point3: do third weight point
calibration, standard weight must be over 10%FS and
be larger than it in CAL.P2, this point can be omitted.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P4
Line calibration point4: do forth weight point
calibration, standard weight must be over 10%FS and
be larger than it in CAL.P3, this point can be omitted.
END.Y
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P5
Line calibration point5: do fifth weight point
calibration, standard weight must be over 10%FS and
be larger than it in CAL.P4, this point can be omitted.
INPUT
Input or view calibration parameters value that is got
before
CAL.END
Calibration end and restart
NOTE:
The details can be referred in section “7.CALIBRATION”