Welch Allyn CP Series Electrocardiographs
Removable USB Storage Devices
CP 100, CP 200
You can now send tests to removable USB storage devices. If both an SD memory card
and a USB storage device are connected, files are sent to the SD card. The capacity of an
SD card or USB storage device must be < 2 GB.
Contact Information for Accessories
CP 100, CP 200
It is now possible to enter contact information that will appear on the accessories Help
page.
To enter or review the contact information, press the Menu key
. Choose
System Settings > Device Configuration > Contact Information.
Spirometry Improvements
CP 200 Spirometry Option
•
Quicker access to spirometry settings
When the electrocardiograph is in spirometry mode, the Menu key
now
immediately opens the Spirometry Settings menu.
•
PCP mode adjustments
In PCP mode, the V/T scaling was changed to AUTO, so the PCP printout now fits on
a single page.
•
Physician's comments page
You can now choose whether spirometry reports include an additional blank page
(with patient information at the top) for physician to write comments on.
To determine whether this page prints with your reports, while in spirometry mode
press the Menu key
. Choose Print Settings > Physician’s Comments > Yes or
No.
•
Correction in Vilajanan profile
The CP 200 Spirometry Option Directions for Use (mat. number 703411 rev C)
indicates that the Vilajanan norm supports FEF25. It does not. It supports FEF75.
•
Protocol settings disabled
When a spirometry protocol is selected (PCP, NIOSH, OSHA, or SSD), its settings are
now disabled in the settings menus. This change avoids confusion during setup.