CAUTION
CAUTION
10
ALARMS
NIBP
ALARM(S)
OFF
ALARM
START /
STOP
SUSPEND/
RESUME
FREEZE/
UNFREEZE
MAIN
MENU
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards (e.g., EN 60950 for data
processing equipment and EN 60601-1 for medical equipment). Furthermore,
all configurations shall comply with the system standard IEC 601-1-1. Anyone
connecting additional equipment to the signal input or output connectors is
configuring a medical system, and is therefore responsible that the system
complies with the requirements of the system standard IEC 601-1-1. If in
doubt, consult your Biomedical Engineering Department.
WARNING
Ensure the network cable is not damaged. A network cable is the sole link
between a Propaq Encore and the Acuity Central Monitor.
4.
Connect the AC adapter to the Propaq Encore and the wall outlet to charge
the battery. Check to see that the battery charging light on the monitor’s
right side panel is on.
5.
Confirm the patient identification at the bedside or enter the patient
information at the Acuity central monitor using the Patient ID Setup
Window.
6.
If alarm limits have not been set, do so at the Propaq Encore or at the
Acuity central monitor using the Alarms Setup Window.
WARNING
If you don’t set alarm limits, Acuity uses preset settings (for arrhythmia limits),
and the last settings saved by the Propaq Encore. If no settings were saved,
Acuity uses the Propaq Encore default settings.
Intl Update Doc.fm5 Page 34 Monday, April 29, 1996 3:30 PM