Replacement parts, Emissions and immunity information – Welch Allyn Image Capture System - User Manual User Manual
Page 38

30
Chapter 5 Maintenance
Welch Allyn Image Capture System
Review Draft
Company Confidential
Aug 24, 2004
Replacement Parts
Emissions and Immunity Information
Canopus ADVC100, Video Converter
700145
Adaptec Fire Connect 4300, PCI
700147
Adaptec Fire Connect 1430, PCMCIA
700146
S-Video Cable
88600
RS-232 Interface Cable
D652020
Table 1.
Guidance and Manufacturer’s Declaration - Emissions
The Image Capture System is intended for use in the electromagnetic environment specified below. The customer
or user of the Image Capture System should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR11
Class B
Group 1
The Image Capture System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR22, FCC
Class B
The Image Capture System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Harmonics
IEC 61000-3-2
N/A
For use in domestically and in Canada.
Beyond the scope of the standard.
Flicker
IEC 61000-3-3
N/A
For use in domestically and in Canada.
Beyond the scope of the standard.
The Image Capture System is suitable for use in all establishments,
including domestic, and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Radiated and Conducted Emissions were tested on the Elgar power source. It complies with the EN55022 Class B
limits, and FCC part 15 which as same limits as EN55011 Group 1, Class B. Harmonics and Flicker testing was not
tested, as the EUT is operated at 120V, 60Hz, which is beyond the scope of those standards.