Appendix B Spirometry Protocols

Welch Allyn CP 200 Spirometry Option

About the NIOSH Protocol

The NIOSH (National Institute for Occupational Safety and Health, U.S.) protocol is for
users who want to make sure that occupational testing and reports meet the
requirements of NIOSH. The device automatically performs as described here, regardless
of user-defined settings.

When using this protocol, the spirometer should be calibrated at three different flows
every day before use.

For details on NIOSH requirements, see the document noted in

Reference 4

on page 75.

When this protocol is selected, testing and reports are affected as follows:

Operational Setting

Adult and Pediatric Norm: NHANES III
(For Asian-Americans the reference equations for Caucasians shall be used, but a
correction factor of 0.94 shall be applied to the predicted values.)

Best Effort Formula: Best Measurement

Composite Norm Values: NO

(For details, see

“Reviewing the Operation Settings”

on page 13.)

Print Settings

Tests: Three Best Efforts
Scale: 20 mm/s & 10 mm/L
Curves: V/T & F/V
Auto Print: YES

(For details, see

“Reviewing the Spirometry Print Settings”

on page 17.)

Calibration Settings

Auto Calibration Report: Yes

(For details, see

“Reviewing the Calibration Settings”

on page 15.)