Chapter 3 Calibrating the Spirometer

Welch Allyn CP 200 Spirometry Option

About Calibration

The American Thoracic Society recommends calibrating a spirometer every day before
testing. In addition, each time you open a new package of flow transducers, verify the lot
number on the package label. If this lot number differs from the lot number used during
the most recent calibration, you must recalibrate the spirometer before resuming testing.

There are two types of calibration:

Single-flow calibration

One inhale/exhale cycle.

Multiple-flow calibration

Three inhale/exhale cycles at three different rates:

3 L in 1 second (3 L/s)
3 L in 3 seconds (1 L/s)
3 L in 6 seconds (0.5 L/s)

For a diagram illustrating this procedure, see

Figure 8

on page 23.

For step-by-step calibration instructions, see

“Performing a Calibration”

on page 24.

For information on reviewing or changing the settings that affect calibration, see

“Reviewing the Calibration Settings”

on page 15.

Note

If you want to add efforts to a saved test, the calibration must stay the same.
Whenever you recalibrate, you lose the ability to add new efforts to tests that
were saved earlier.

Caution For proper performance, the calibration syringe itself must be
recalibrated every year. See its calibration certificate for the most recent
calibration date. When the syringe is due for recalibration, return it to the
manufacturer. For details, see

“Service Policy”

on page 51.