Directions for Use

Chapter 4 Performing Spirometry Tests

39

Recording a Test

To record a spirometry test, follow these steps.

1.

Measure the patient’s standing height to the nearest half inch (or centimeter) in
stocking feet.

Accuracy is important; height greatly influences the predicted values.

2.

If the patient’s demographics do not match the current spirometry norm, select a
more appropriate norm.

To find out how, see

“Select Adult Predictive Norm”

on page 13 or

“Select Ped.

Predictive Norm”

on page 13.

3.

Press .

The first time this key is pressed each day, the prompt “Calibrate Now?” appears.

4.

(Optional) Calibrate. See

“Calibrating the Spirometer”

on page 21.

The following screen appears.

Figure 27. “Enter New Patient” Screen

Note

If the patient has obvious spinal deformities, measure the arm span from
fingertip to fingertip with arms outstretched against a wall. Enter the arm
span instead of height.

To learn how to choose which fields display
here, see

“Reviewing the Patient Data Fields

Available”

on page 19.

Enter New Patient

Search

Schedule

Clear

Done

9:17AM Oct 16 08

Patient ID
Last Name
Birth Date
Height
Gender
Race
Smoke Years

/

/

MM / DD / YYYY

ft.

in.

yr.

Use up and down arrows to change fields