B) parts counting, Parts counting – Torbal AD50 Series Manual User Manual
Page 22

Instruction Manual - TORBAL AD Series
(B) Parts Counting
1. To enable parts counting press the “T” (YES) key when Parts is displayed.
2. Commands Co On and Co OFF will be displayed sequentially. To enable parts counting and
proceed with configuration press the “T” (YES) key when Co On is displayed. If the function
is already enabled and you wish to disable it, press the “T” (Yes) key when Co OFF is
displayed.
22
Table 7.1.3.A: Animal Weighing Operating Modes
Automatic
Weighing Initialization: Automatic
Result Clearing: Automatic
Tare: Automatic
When set to Automatic, the scale will automatically
begin the animal weighing process. Once the animal
is removed from the weighing pan, the scale will
automatically tare and prepare for the next weighing.
Semi-Automatic
Weighing Initialization: Automatic
Result Clearing: Manual
Tare: Manual
When set to Semi-Automatic, the scale will
automatically begin the animal weighing process.
Once the animal is removed from the weighing pan
the result remains displayed until it is manually
cleared. The scale must be manually tared before the
next weighing.
Manual
Weighing Initialization: Manual
Result Clearing: Manual
Tare: Manual
When set to Manual, the animal weighing process
must be manually initialized. Once the animal has
been weighed and the pan cleared the result must be
manually cleared. The scale must be manually tared
before the next weighing.