7.1.1.2. Start
To calibrate the scale, follow the steps below:
1. Press “M” to enter the Main Menu
2. When CALIB is displayed, press “T” (YES).
3. The scale will display commands Start and Setup sequentially. Press , “T” , when Start is displayed
.
4. The scale will display CAL, indicating that the calibration mode has been entered.
5. Before calibration begins, the scale will perform automatic tare. Taring will be displayed during auto
tare.