11 specifications – Rice Lake IQ Plus 355 Installation Manual V1.17 User Manual
Page 50

46
IQ plus 355 Installation Manual
7.11
Specifications
Power
Line Voltages
115 or 230 VAC
Frequency
50 or 60 Hz
Power Consumption
70 mA @ 115 VAC (8W)
35 mA @ 230 VAC (8W)
Fusing
115 VAC
2 x 200 mA TR5 subminiature fuses
Wickmann Time-Lag 19374 Series
UL Listed, CSA Certified
230 VAC
2 x 100 mA TR5 subminiature fuses
Wickmann Time-Lag 19372 Series
UL Recognized, VDE Approved
Analog Specifications
Full Scale Input Signal Up to 45 mV
Excitation Voltage
10 ± 0.5 VDC,
8 x 350
or 16 x 700 load cells
Sense Amplifier
Differential amplifier with
4- and 6-wire sensing
Analog Signal
Input Range
0.6 mV/V – 4.5 mV/V
Analog Signal
Sensitivity
0.3
V/graduation minimum,
1.5 mV/grad recommended
Input Impedance
200 M
, typical
Noise (ref to input)
0.3
V p-p with digital filters at 4-4-4
Internal Resolution
308,000 counts
Display Resolution
100,000 dd
Measurement Rate
Up to 30 measurements/sec
Input Sensitivity
150 nV per internal count
System Linearity
Within 0.01% of full scale
Zero Stability
150 nV/°C, maximum
Span Stability
3.5 ppm/°C, maximum
Calibration Method
Software, constants stored in
EEPROM
Common Mode
Voltage
–0.15 to +0.95 V, referred to earth
Common Mode
Rejection
130 dB minimum @ 50 or 60 Hz
Normal Mode
Rejection
90 dB minimum @ 50 or 60 Hz
Input Overload
± 12 V continuous, static discharge
protected
RFI Protection
Signal, excitation, and sense lines
protected by capacitor bypass
Analog Output
Optional: fully isolated, voltage or
current output,14-bit resolution.
Voltage output: 0 –10 VDC
Load resistance:1K
minimum
Current output: 4–20 mA
External loop resistance: 500
maximum
Digital Specifications
Microcomputer
Phillips 80C51XA-G49 main processor
@ 9.8304 MHz
Digital Inputs
2 inputs, TTL or switch closure,
active-low
Digital Filters
3 filters, software selectable
Serial Communications
EDP Port
Full duplex RS-232
Printer Port
Output-only RS-232 or active 20 mA
current loop
Both Ports
9600, 4800, 2400, 1200, 600, 300
bps; 7 or 8 data bits; even, odd, or no
parity
Operator Interface
Display
6-digit LED display. 7-segment, 0.8 in
(20 mm) digits
LED annunciators
Gross, net, center of zero, standstill, lb/
primary units, kg/secondary units, oz, g
Keyboard
5-key flat membrane panel
Environmental
Operating Temperature –10 to +40°C (legal);
–10 to +50°C (industrial)
Storage Temperature
–25 to +70°C
Humidity
0–95% relative humidity
Enclosure
Enclosure Dimensions 9.5 in x 6 in x 2.75 in
24 cm x 15 cm x 7 cm
Weight
2.8 Kg (6.1 lb)
Rating/Material
UL Type 4X
Certifications and Approvals
NTEP
CoC Number
97-130A1
Accuracy Class
III/III L
n
max
: 10 000
Measurement Canada
Approval
AM-5213
Accuracy Class
III/III HD
n
max
: 10 000
OIML
R76-2 Test Certificate TC6008
Approval T5692
Accuracy Class
III
n
max
: 3 000
s
.
!
4
)/
. !
, # / . & % 2
%
.
#
%
s
/
.
7
%
) '
( 4
3 ! . $ -
% !
3
5
2
%
3