0 appendix, 1 error messages, 1 displayed error messages – Rice Lake IQ Plus 355 Installation Manual V1.17 User Manual
Page 39: Appendix, Displayed error messages

Appendix
35
7.0
Appendix
7.1
Error Messages
The IQ plus 355 indicator provides a number of error messages. When an error occurs, the message is shown on
the indicator LED display. Error conditions can also be checked remotely by using the XE EDP command as
described in Section 7.1.2.
7.1.1
Displayed Error Messages
The IQ plus 355 provides a number of front panel error messages to assist in problem diagnosis. Table 7-1 lists
these messages and their meanings.
Error Message
Description
Solution
E A/D
A/D physical error
Call Rice Lake Weighing Systems (RLWS) Service.
EEEROM
EEPROM physical error
EVIREE
Virgin EEPROM
Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.
EPCKSM
Parameter checksum error
EACKSM
A/D calibration checksum error
A/D converter requires recalibration. Call RLWS Service.
EFCKSM
Printer format checksum error
Call RLWS Service.
ELCKSM
Load cell calibration checksum error
Recalibrate load cells.
EIDATA
Internal RAM checksum error
Call RLWS Service.
E REF
A/D reference error
A/D converter requires recalibration. Call RLWS Service.
Error
Internal program error
Check configuration. Run XE command (see Section 7.1.2 on
page 36) to determine error type.
Call RLWS Service if unable to clear error by cycling power or if error
recurs.
OVERFL
Overflow error
Weight value too large to be displayed.
- - - - - -
Gross > overload limit
Gross value exceeds overload limit. Check configuration or signal
input level. Overload can be caused by input signal > 45mV or
common mode voltage > 950 mV.
_ _ _ _ _ _
A/D underrange
A/D reading < –4 mV. Check scale for binding or damage.
Table 7-1. IQ plus 355 Error Messages