IQ plus 355 Installation Manual
6.1.2
Reporting Commands
Reporting commands (see Table 6-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.
6.1.3
The RESETCONFIGURATION Command
The RESETCONFIGURATION command can be
used to restore all configuration parameters to their
default values. Before issuing this command, the
indicator must be placed in test mode (press and hold
setup switch for approximately three seconds to show
TEST menu).
This command is equivalent to using the DEFLT
function on the TEST menu. See Section 7.9 on
page 44 for more information about test mode and
using the TEST menu.
NOTE:
All load cell calibration
settings are lost when the RESETCONFIGURATION
command is run.
6.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 6-3 through 6-10).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command
Most parameter values can be changed in setup mode
only. Use the following command syntax when
changing parameter values:
command=value
where
value
is a number or a parameter value. Use no
spaces before or after the equal (=) sign. If you type an
incorrect command or value, the display reads
??
.
Changes to the parameters are saved as they are
entered but typically do not take effect until you exit
setup mode.
For example, to set the motion band parameter to 5,
type the following:
MOTBAND=5D
Command
Function
DUMPALL
List all parameter values
VERSION
Write IQ plus 355 software version
P
Write current displayed weight with units
identifier. See Section 7.2 on page 36 for
more information.
ZZ
Write current weight and annunciator
status. See Section 7.2 on page 36 for
more information.
S
Write one frame of stream format
Table 6-2. EDP Reporting Commands
Command
Description
Values
GRADS
Graduations
1–100 000
ZTRKBND
Zero track band
OFF, 0.5D, 1D, 3D
ZRANGE
Zero range
1.9%, 100%
MOTBAND
Motion band
1D, 2D, 3D, 5D, 10D, 20D, OFF
OVRLOAD
Overload
FS+2%, FS+1D, FS+9D, FS
SMPRAT
Sample rate
15HZ, 30HZ, 7.5HZ, 3.75HZ
DIGFLTR1
DIGFLTR2
DIGFLTR3
Digital filtering
1, 2, 4, 8, 16, 32, 64
DFSENS
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
DFTHRH
Digital filter cutout threshold
NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD
TAREFN
Tare function
BOTH, NOTARE, PBTARE, KEYED
Table 6-3. CONFIG EDP Commands