2 scale capacity, 3 load cell mv, 4 total load cell build – Rice Lake BCi Belt Scale - Installation & Operation Manual Version 2.00 User Manual
Page 46: 5 rate unit time, 6 filter (s), 7 filter threshold (divisions), 8 speed unit time, 9 fixed speed, 10 unit of measure, 11 unit of rate

40
BCi
Installation & Operation Manual
There are many parameters in the 
Supervisor Mode (shown in the following 
sections), but most will not be used for a 
basic installation. The parameters that 
must be set prior to calibration include the following:
Calibration Weight (if using static weights to calibrate)
•
Idler Spacing
•
Number of Idlers
•
Belt Length
•
Load Cell MV
•
Total Load Cell Build
The other parameters that are commonly set are the 
filtering values but they can be done after setup and 
calibration.
The angle of inclination is not required with 
Rice Lake Weighing Systems frames due to 
laterally opposed shearbeams.
4.2
Scale Capacity
This parameter defines the maximum rated capacity 
for the belt scale or the maximum amount of weight 
that can pass along the belt scale per hour. 
The default parameter is measured in tons/hour.
Enter the maximum rate in tons using the numeric 
keypad and press the 
enter
key to save that parameter.
4.3
Load Cell MV
Enter the average mv/v rating of all the load cells in 
the system. 
Use the numeric keypad to enter that value which 
must be greater than 0. Press the 
enter
key to save that
parameter.
4.4
Total Load Cell Build
This parameter is the total load cell capacity of the all 
the load cells in system. If you have eight load cells 
for example and each one has a rated capacity of XX, 
take that capacity and times it by 8 to get the total load 
cell capacity to be entered into the BCi integrator. 
4.5
Rate Unit Time
This parameter defines what unit of time the rate will 
be displayed in on the BCi. Available selections are:
•
Hr. - hour (default)
•
Min - minute
•
Sec - seconds
Press the
enter
key to toggle between the listed
choices above. The default rate unit time is Hr.
4.6
Filter (s)
This parameter defines the number of seconds 
filtering samples to average the scale load. It is 
measured in tenths of seconds.
4.7
Filter Threshold (divisions)
This is the number os division the load must maintain 
for the filter to be enabled. If the load sees a change in 
weight that is greater than this value, the filter is 
flushed. 
4.8
Speed Unit Time
This parameter defines how the belt speed is displayed 
on the BCi. The selections are displayed in the 
following units.
•
Hr. - hour
•
Min - minute (default)
•
Sec - seconds
Press the
enter
key to toggle between the listed
choices above. The default speed unit time is 
displayed in minutes.
4.9
Fixed Speed
This parameter allows a technician to configure a 
fixed belt speed for the BCi. The BCi will no longer 
calculate speed using the speed sensor. 
The default value is 0 and if a value is set then the BCi 
will used the fixed value. The unit of measure is how 
the BCi has been configured for speed so usually it 
will be either ft/min or m/min. 
The user can change the time units to be either 
seconds or hours. Enter a numeric value and press the 
enter
key to save that value.
4.10 Unit of Measure
The unit of measure parameter defines how the belt is 
measured and speed calculated. The selections are 
displayed in the following units.
•
Ft - feet (default)
•
M - meters
Press the
enter
key to toggle between the listed
choices above. The default speed is shown as Ft/Min.
4.11 Unit of Rate
The unit of rate parameter defines how the rate is 
displayed. The selections are displayed in the 
following units: 
•
TN (tons) - (default)
•
lb (pounds)
•
kg (kilograms
Note