Rice Lake 480 Legend Series Digital Weight Indicator Installation Manual User Manual
Page 31

Configuration 27
3.3.1
Configuration Menu Structures and Parameter Descriptions
Figure 3-10. Configuration Menu
CONFIG Menu
Parameter
Choices
Description
GRADS
10000
1 – 100000
Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1 – 100 000 and should be consistent with legal requirements and
environmental limits on system resolution.
ZTRKBN
0 dd
0.0 – 3.0dd
Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZRANGE parameter). Selections are ±
display divisions. Maximum legal value varies depending on local regulations.
ZRANGE
2%
1% – 100%
Zero range. Selects the range within which the scale can be zeroed. The 2% selection is ±
2% around the calibrated zero point, for a total range of 4%. Indicator must be at
standstill to zero the scale.
INIZR
0 %
0 – 100%
Initial ZERO Range at power up. See Error messages HINOFF and LINOFF in
Section 9.1.1
If on power up, the weight value is between the ±% range specified of Calibrated Zero,
the indicator will zero off that weight. If outside that range, the indicator will display
HINOFF or LINOFF, depending on whether the initial reading is above or below the range.
If not set at zero, hopper/scale MUST be empty when powered up.
MOTBAN
0.1D
0.1D – 100D
Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected, the standstill symbol lights.
• Motion window is the consecutive A/D updates compared to the RD-MOT settings.
• Stability window is the consecutive A/D updates compared to the RD-STA settings.
Some operations, including print, tare, and zero, require the scale to be at standstill.
Maximum legal value varies depending on local regulations.
RD-STA
2
1 – 20
Read Stability – Defines the number of sequential samples within the motion band for
which a reading is regarded as stable; defined in conjunction with MOTBAND and
RD-MOT parameters. When the applied weight has settled within the stability range, a
rolling averaging filter is applied to improve the stability.
Table 3-7. Configuration Menu Parameters
GRADS
010000
CONFIG
FORMAT
CALIBR
PROGRM
DIG IN
ALGOUT
PASWRD
TEST
MENU
0 dd
ZTRKBN
ZRANGE
2 %
INIZR
0 %
MOTBAN
1 dd
RD-STA
002
RD-MOT
001
OVRLOA
FS-2
FS-1D
FS-9D
FS
DSPRAT
0.10SEC
0.25SEC
0.50SEC
0.75SEC
1.00SEC
NONE
SMPRAT
5 HZ
10 HZ
20 HZ
40 HZ
DIGFIL
ON
OFF
DFSENS
LIGHT
MEDIUM
HEAVY
DFTHRH
000000
TAREFN
BOTH
NOTARE
PBTARE
KEYED
Note