1 seal the indicator, Seal the indicator, Allows service access prevents service access – Rice Lake 480 Legend Series Digital Weight Indicator Installation Manual User Manual
Page 19

Installation 15
2.6.1
Seal the Indicator
For access to configuration parameters, the setup switch must be pressed.
Allows Service Access
The calibration cover is used for inserting a lead wire seal through both fillister screws and the plastic cover. The
cover allows access to the electronics and electrical contacts, while preventing access into Legal for Trade
configuration parameters.
There is an alternative cover over the A/D to ensure the A/D and load cell connection cannot be changed after the
unit is sealed.
There is also an audit trail counter to track calibration and configuration changes made to legally relevant
parameters.
Place the Calibration Cover on the backplate and secure with the existing fillister head screws and o-rings to seal
the indicator for Legal for Trade approval.
Figure 2-6. Sealing the Indicator
Prevents Service Access
The calibration cover is used for inserting a lead wire seal through 3 fillister screws. This prevents access to the
electronics, electrical contacts, and Legal for Trade configuration parameters.
Note
Allows Service Access
Prevents Service Access
With A/D cover
12 - Screw, Pan Cross Head
8-32NC
Calibration Cover
for Hardware Sealing
Setup Switch
Access
1 - Screw, 4-40 drilled
fillister head
1 - Screw, 8-32 drilled
fillister head
Without A/D cover
11 - Screw, Pan Cross Head
8-32NC
1 - Screw, 8-32 drilled
fillister head
Calibration Cover
for Hardware Sealing
2 - Screw, 8-32 drilled
fillister head