Model 7i Digital Force / Torque Indicator

User’s Guide


34

Press ENTER.

11. At the completion of the tension calibration, the display appears as follows:

To save the calibration information, select “Save & exit”. To exit without saving the data select
“Exit without saving”.

12. Any errors are reported by the following screens:

Displayed at the start of calibration if a disallowed unit is selected.

Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.

CALIBRATION

Load not stable.

Please try again.

CALIBRATION

Units must be gF.

Please try again

Press ENTER.

CALIBRATION

COMPRESSION

Load too low.

Please try again.

CALIBRATION

COMPLETE


Save & exit

Exit w/o saving

CALIBRATION

COMPRESSION COMPLETE

Reverse direction
for tension.
Attach necessary
weight fixtures,
then press ENTER.