Model 7i Digital Force / Torque Indicator

User’s Guide


32

Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return
to the menu without saving changes.

3. After the number of calibration points has been entered, press ENTER. The display appears as

follows:

4. Place the force sensor horizontally on a level surface free from vibration, then press ZERO. The

indicator will calculate internal offsets, and the display appears as follows:

If failed:

5. The following screen appears after the offsets have been calculated:

Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.

6. The display appears as follows:

CALIBRATION

NOT COMPLETE


Cancel

Exit w/o saving

CALIBRATION

COMPRESSION

Attach necessary
weight fixtures,
then press ENTER.

CALIBRATION

OFFSET

Sensor failed

Analog failed

CALIBRATION

OFFSET

Sensor passed
Analog passed

CALIBRATION

OFFSET


Please wait…

CALIBRATION

OFFSET

Place sensor
horizontally, then
press ZERO.