Cooper Instruments & Systems Model 7i Professional Force/Torque Indicator User Manual
Page 33
![background image](/manuals/600693/33/background.png)
Model 7i Digital Force / Torque Indicator
User’s Guide
32
Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return
to the menu without saving changes.
3. After the number of calibration points has been entered, press ENTER. The display appears as
follows:
4. Place the force sensor horizontally on a level surface free from vibration, then press ZERO. The
indicator will calculate internal offsets, and the display appears as follows:
If failed:
5. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.
6. The display appears as follows:
CALIBRATION
NOT COMPLETE
Cancel
Exit w/o saving
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures,
then press ENTER.
CALIBRATION
OFFSET
Sensor failed
Analog failed
CALIBRATION
OFFSET
Sensor passed
Analog passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place sensor
horizontally, then
press ZERO.